Troubleshooting 6-1
6. Troubleshooting
Problem What should I do? Remarks
1.
“Disconnect” error
message appears on the
“scanner interface”
screen
1. Check:
Power cable connected?
USB cable connected?
2. Turn system off; then on again.
2.
Onyx-Rad software
does not load. Error
message received
Check if Dongle is connected to the
parallel port of the PC
See also: Chapter 4:
Installation
3.
GOP filters do not
apply on the image
Make sure that the GOP filters are
licensed
See also: Chapter 4:
Installation
4.
After the system is
turned on, initialization
sound does not stop.
Turn system off; then on again.
If the problem persists, contact your
distributor.
5.
Cassette cannot be
inserted into the tray.
Make sure the cassette is positioned at the
center of the tray.
6.
Image is too bright. From the Scanner Interface – Setup menu,
increase “PM-Gain” for the type of image
which is too bright
See also: User Manual –
Chapter: Using the
scanner interface
7.
Image is too dark From the Scanner Interface – Setup menu,
and decrease “PM-Gain” for the type of
image which is too light
See also: User Manual –
Chapter: Using the
scanner interface
8.
Double image received 1. Check that the erase lamps are
working
2. If the erase lamps are OK, increase
the erasing factor by 5 from the
Scanner Interface – Setup menu Î
Configuration
9.
White lines appear on
image.
Contact Orex service department
10.
Horizontal lines appear
on image
Contact Orex service department
11.
Scan does not start Contact Orex service department
12.
Systems is shaking Place system on steady surface
13.
Diagonal lines appear
on image
There is light penetration into the system.
Place the system away from intense light
source (window).
See also Chapter 3: Site
Preparation